DBBC2 calibration

    Table of contents

    Version as of 12:20, 9 Nov 2024

    to this version.

    Return to Version archive.

    View current version

    - Everytime a new DBC firmware version is installed (PFB or tunable) the calibration parameters must be determined. -

    -  Calibration should also be repeated and checked whenever the Control Software has been restarted (especially when using the DDC mode)


    Calibration Procedure


    1) Inject a noise signal into  all of the DBBC boards (e.g. use a splitter)

    2) Start the DBBC control software  (PFB or DDC) and wait until programming of the boards is finished

    3) Adjust the IF power levels to around 10000

    • PFB and DDC mode

    In the DBBC2 client manually set the attenuation (e.g. 40) until you reach a target of 10000 counts.

    dbbcifa=1,40,1
    

    Check the counts with:

    dbbcifa
    

    Change attenuation and repeat until the target has been reached. Repeat for all boards.

    • DDC mode only
    dbbcgain=all,20,20
    

    4) start the calibration

    •  PFB mode:
    calibration
    
    • DDC mode:
    calibration=all
    

    5) Update the configuration file

    A file containing the results is produced in c:\ (e.g. DATE-calib.txt) by the calibration command. Inspect or plot the DATE-calib.txt file and search for the minimum between the two peaks.Insert the minimum values into the configuration file e.g. c:\DBBC_CONF\dbbc_config_file_104.txt (file name varies depending on the version). Edit the line just above CAT2 entry.