    3.4.2013

    Problem with 160 MHz mode

    When trying to do calibration observations in rcumode 6, the RSPDriver kept failing when setting the station clock to 160 MHz. From looking at the logfiles of the RSPDriver it seems to be a problem with the Clock- (TDS-) boards. When taking subrack 5 (in which the TDS board was replaced in March) out of the station config, then the system could be set to the 160 MHz mode.


    Monitoring software installed

    The iStnMonitor software has been installed. The plots can be viewed at: https://lofarx.mpifr-bonn.mpg.de/cacti/graph_view.php

    The LOFAR Environment will be expanded when new sensors for the container monitoring will become available.

     

    2013 Apr 08

    Bad (oscillating?) tiles in HBA 5 mode on startup in local mode:

    RCUs 18,19,34 are very bad

    RCUs 50,51,52 are modestly bad (high ringing around line at 161 MHz)

    Tiles 9,17 are very bad

    Tiles 25, 26 are modestly bad

    These tiles are neighbors.  See tiles.0.pdf

     

    RCU 82 (Tile 41) also has low-level ringing

     

    Set pulsar RCU command to 0:17,20:33,36:49,54:81,84:191

     

    With those tiles turned off, there is no more 161 MHz line.

     

    When the bad tiles are in, the spectra from RCUs look like HBA_5_20130408_bad.jpg and when they are not powered up, the spectra of the remaining RCUs look like HBA_5_20130408_good.jpg

     

    James M Anderson

    9.4.2013

    Oscillating tile(s)

    The were no strong oscillations today, as there were yesterday. But tile 9 (RCUs 18, 19) showed a few "bumps" in the spectrum. These went away when FE-module 16 (the last one) was switched off.

    So tile 9, FE 16 needs to be replaced.