Continuous calibration

    Version as of 10:13, 29 Mar 2024

    to this version.

    Return to Version archive.

    View current version

    Table with the current status of the EVN stations for the continuous calibration (band dependent)

     

    Current status of the continuous calibration implementation at the stations

     
      L Band C band M band X band K band Q band Last updated
    Ar              
    Bd              
    Ef              
    Hh              
    Jb              
    KVN              
    Mc              
    Nt              
    On              
    Sr              
    T6              
    Tr              
    Sv              
    Wb              
    Ys - 80 Hz 80 Hz 80 Hz 80 Hz No 5-06-2015
    Zc              

     

    • The DBBC allows to detect synchronously an 80 Hz signal integrating the power from the cal off and cal on phases.
    • The 80 Hz signal can be obtained from output "Cont Cal".
    • Continuous cal can be enabled by typing:
      • cont_cal=on (at the FS)
    • In some cases the signal is inverted and you can get negative Tsys. This is a pending issue to be solved. It can be solved by patching the FS, but this solution should be avoided in the future and look for a cleaner one. The DBBC allows to invert the signal but this does not work at V105 version yet.
    • station.prc should be ready to allow have continuous calibration fro some bands and not for others.