Table of contents
Combined revision comparison
...
- Start the Field System.
- Check times with FMSET.
- Check the correct firmware is loaded to the DBBC and Fila10G. (dbbc=version, fila10g=version)
- Check recorder (mk5=version?)
- Check the correct clock and input settings are set on the Fila10G (fila10g=sysstat). Might needs loading of the setup procedure before (e.g. proc=c182aef, setup01, ... wait and check)
- After loading the setup procedure:
- Check input level on IF
- Check counts on PFB channels.
- Check system temperatures with caltsys (if possible)
- Check correct synthesizer frequency. If possible, inject a tone and check on spectrum analyser if it appears at the correct frequency.
- Turn on phasecal for recording tests
- Do a short test recording and run the checkdata.py script to look for proper bit statisitics and channel bandpass (see here as well). Phasecal tones can be checked here as well.
)
Local test with antenna
- Adjust pointing and focus
- If possible, observe a maser or line source to confirm correct frequency settings.
Version from 11:40, 22 Mar 2019
...
- Start the Field System.
- Check times with FMSET.
- Check the correct firmware is loaded to the DBBC and Fila10G. (dbbc=version, fila10g=version)
- Check the correct clock and input settings are set on the Fila10G (fila10g=sysstat). Might needs loading of the setup procedure before (e.g. proc=c182aef, setup01, ... wait and check)
- After loading the setup procedure:
- Check input level on IF
- Check counts on PFB channels.
- Check system temperatures with caltsys (if possible)
- Do a short test recording and run the checkdata.py script to look for proper bit statisitics and channel bandpass (see here as well
)
Current version
...
- Start the Field System.
- Check times with FMSET.
- Check the correct firmware is loaded to the DBBC and Fila10G. (dbbc=version, fila10g=version)
- Check recorder (mk5=version?)
- Check the correct clock and input settings are set on the Fila10G (fila10g=sysstat). Might needs loading of the setup procedure before (e.g. proc=c182aef, setup01, ... wait and check)
- After loading the setup procedure:
- Check input level on IF
- Check counts on PFB channels.
- Check system temperatures with caltsys (if possible)
- Check correct synthesizer frequency. If possible, inject a tone and check on spectrum analyser if it appears at the correct frequency.
- Turn on phasecal for recording tests
- Do a short test recording and run the checkdata.py script to look for proper bit statisitics and channel bandpass (see here as well). Phasecal tones can be checked here as well.
Local test with antenna
- Adjust pointing and focus
- If possible, observe a maser or line source to confirm correct frequency settings.