Table of contents
You are currently comparing two old versions - only when you are comparing against the latest version can you revert. Return to version archive.
Combined revision comparison
...
- close any other running control software programs
- close the DBBC3 client program
- double click the icon labeled "DBBC3 Control OCT_D_
OCT_D_v110.exev120.exe" "(Version number can be different) - answer first question with "y" in order to do a full reload of the firmware.
- wait until the control software has fully loaded and responds with "Waiting for connection on port 4000"
ValidateLoad the filters and validate the system
- make sure the DBBC3 client is not running
- start the dbbc3ctl program (on the EHT Control Computer cc-pico):
cd /home/oper/rottmann/dbbc3/utilities ./dbbc3ctl.py dbbc3setupDBBC3_OCT_D.py dbbc3 -n 4
- do a full system check:a
number of tests and verifications the script will ask whether you want to load the
in dbbc3ctl.py :filters. Answer with 'y'.
check systemNote: any checks prior to this question will fail if the filters have already been loaded.all
...
Other changes:
- /body/pre[3]/@style: nothing ⇒ "margin-left: 40px;"
Version from 15:19, 11 Jan 2022
...
- close any other running control software programs
- close the DBBC3 client program
- double click the icon labeled "DBBC3 Control
OCT_D_v110.exe"(Version number can be different) - answer first question with "y" in order to do a full reload of the firmware.
- wait until the control software has fully loaded and responds with "Waiting for connection on port 4000"
Load the filters and validate the system
...
cd /home/oper/rottmann/dbbc3/utilities ./setupDBBC3_OCT_D.py dbbc3 -n 4
After carrying out a number of tests and verifications the script will ask whether you want to load the filters. Answer with 'y'.
Note: any checks prior to this question will fail if the filters have already been loaded.
...
Version as of 15:24, 11 Jan 2022
...
- close any other running control software programs
- close the DBBC3 client program
- double click the icon labeled "DBBC3 Control OCT_D_v120.exe"
- answer first question with "y" in order to do a full reload of the firmware.
- wait until the control software has fully loaded and responds with "Waiting for connection on port 4000"
Validate the system
- make sure the DBBC3 client is not running
- start the dbbc3ctl program (on the EHT Control Computer cc-pico):
cd /home/oper/rottmann/dbbc3/utilities ./dbbc3ctl.py dbbc3
- do a full system check:
in dbbc3ctl.py :
check system all
...